SR EN IEC 60942:2018
Electroacustică. Calibratoare acustice
This document specifies the performance requirements for three classes of sound calibrator: class LS (Laboratory Standard), class 1 and class 2. Acceptance limits are smallest for class LS and greatest for class 2 instruments. Class LS sound calibrators are normally used only in the laboratory; class 1 and class 2 are considered as sound calibrators for field use. A class 1 sound calibrator is primarily intended for use with a class 1 sound level meter and a class 2 sound calibrator primarily with a class 2 sound level meter, as specified in IEC 61672-1. The acceptance limits for class LS sound calibrators are based on the use of a laboratory standard microphone, as specified in IEC 61094-1, for demonstrations of conformance to the requirements of this document. The acceptance limits for class 1 and class 2 sound calibrators are based on the use of a working standard microphone, as specified in IEC 61094-4, for demonstrations of conformance to the requirements of this document. To promote consistency of testing of sound calibrators and ease of use, this document contains three
Status :
In vigoare
Data aprobării : 28.09.2018
Număr de pagini : 119
ICS : 17.140.50 Electroacustica,33.100.20 Imunitate
Comitet tehnic : 26 - Electroacustica şi ultrasunete
- Inlocuieste SR EN 60942:2004