SR EN 50527-1:2017

Procedură pentru evaluarea expunerii la câmpuri electromagnetice a lucrătorilor purtători de dispozitive medicale implantabile active. Partea 1: Generalităţi

This European Standard provides a procedure to assess the risk to workers bearing one or more active implantable medical devices from exposure to electric, magnetic and electromagnetic fields at a workplace. It describes how a general risk assessment should be performed and determines whether it is necessary to carry out a detailed risk assessment. NOTE 1 This European Standard does not cover indirect effects caused by non active implants. NOTE 2 The risk of human exposure to EMF considered is only due to malfunctioning of AIMD. Possibilities of AIMD contribution to the risk, e.g. local modification of the distribution of EMF produced by external source or production of own EMF, are covered by the respective product standards for the AIMD. Based on specific workplace standards it can be determined whether preventive measures/actions need to be taken to comply with the provisions of Directive 2013/35/EU. The work situation covered is considered to be under normal working conditions including normal operation, maintenance, cleaning and other situations being part of the normal w

190,15 Lei

Status : In vigoare
Data aprobării : 30.06.2017
Număr de pagini : 35
ICS : 11.040.40 Implanturi chirurgicale, proteze si orteze,13.100 Securitate profesionala. Igiena industriala,13.280 Protectia contra radiatiilor
Comitet tehnic : 279 - Expunerea corpului uman la câmpuri electromagnetice

Relaţii cu alte standarde: